Contact an Account ManagerStandard SPE Procedure for CHROMABOND® HR-P
Summary
Type: SPE
Application #: 304034
Catalog#: 730108 - SPE Cartridges, Chromabond, HR-P, 3ml, PP, 200mg 30/pk
Detection:
Conditions: Column conditioning: 3 mL methanol, then 3 mL water.
(Do not let run the column dry.)
Sample aspiration: The prepared sample is passed through the column by vacuum or pressure (max. 1000 mL sample volume)
Washing: With 3 mL methanol - water (2:98, v/v) are washed impurities
Elution: Analytes are eluted with 2 x 2 mL methanol/acetone (1:1, v/v) or dichloromethane
Authors: SPE department
Source: Macherey-Nagel